Troubleshooting and Test Section
Problem
Power switch does not
illuminate in the ON
state
Confidence test
Video is not displayed
8
Video is not displayed
from a given output port
External control failure
17
Possible Causes and Remedies
Loose power cable: Verify that the power cable is fully inserted into the unit's
power connector and that the power supply's connector collar is properly
threaded onto the ScaleOR's power connector.
Defective power supply: Replace the power supply.
Wall socket: Some wall sockets have on / off switches built in. If the socket you
are using has a built in switch, verify that is in the on position.
Mitigation:
Bypass the ScaleOR by connecting the output of the video source directly to a
monitor.
Turn the ScaleOR off, then connect its
monitor. Turn the
ScaleOR
approximately 10 seconds after the ScaleOR is turned on. Turn the ScaleOR off,
then connect its
SDI
output port to the monitor and repeat the test. If your
ScaleOR has a
Fiber Optic
Optic
output port to the monitor and repeat the test.
If the ScaleOR's power button's blue ring is flashing once per second, then the
unit is not detecting a video signal. Verify that an input signal is connected to
the unit's input. If there is a connection try replacing the video cable. If video
appears, then the cable is bad. If video does not appear, then connect the unit
to a different video source. If video appears, then the previous video source is
turned off, or defective. If video still does not appear, then the ScaleOR's Input
module may be defective, or not properly seated. Try reseating the input
module. If that doesn't clear the problem then the input module is defective.
If you determine that the input module is defective, please contact customer
service for assistance.
Mitigation:
If possible, connect the video source directly to the video display.
Verify that video is displayed from the
DVI
port is displayed, and not from the
may set to a
DVI
only mode (see
connecting the
SDI
port to a monitor may be bad or the
If video from the
SDI
port is displayed, and not from the
from the
DVI
port may be bad, or the
If you determine that one of the output ports is defective, please contact
customer service for assistance.
USB cable is loose or defective: Verify that the USB cable is fully inserted into the
ScaleOR's USB port and the PC's USB port.
Control program has failed: Restart the program
Controlling PC has failed: Restart the PC
Mitigation:
Control the ScaleOR from its front panel Keypad
DVI
output port to a powered up
on, the NDS logo should appear on the monitor
output, turn the ScaleOR off, then connect its
DVI
or
SDI
output port. If video from the
SDI
port then ScaleOR's output mode
Output Modes Table
DVI
port is defective.
fiber
on page 15), the cable
SDI
output is defective.
DVI
port then the cable