EN
sleePCUbe CleaninG
WARNING
To avoid electric shock, always unplug the power cord from the wall outlet power source before cleaning.
CAuTION– Never rinse or place the device in water. Never allow liquids to get into or around any of the ports, switches or air filter; doing so will result in
product damage. If this occurs, do not use the device. Contact your home care provider for service.
Device Cabinet
1.
Unplug the SleepCube from the power source. Wipe the enclosure with a clean, damp cloth every few days to keep the enclosure dust free.
2.
Allow the device to dry completely before plugging it into a power source.
Filters
Air-Inlet Filter (Standard)
1.
The air-inlet filter should be checked every 10 days and cleaned as necessary. The filter is located in the rear of the device.
CAuTION– Proper filter function is important for the operation of the device and to protect the device from damage.
2.
To clean, remove the dark outer foam filter from the frame.
3.
Wash the filter in a solution of warm water and dishwashing detergent, and rinse with water. Allow the filter to dry completely. If the filter
(part #DV51D-602) is damaged, contact your home care provider for a replacement.
4.
Ensure the filter is completely dry before installation and use.
Fine Particle Filter (Optional)
1.
If you have an optional fine particle filter (part #DV51D-603), inspect this filter and replace it if it shows signs of dirt or damage. Do not wash.
2.
Install the fine particle filter behind the standard filter.
NOTE
– If the fine particle filter is not installed first according to the directions, the life of the filter will be reduced requiring more frequent replacement.
Tubing
The air supply tubing should be cleaned daily. Remove the tubing from the device and mask. Use a mild detergent and water to clean the inside of
the tubing. Rinse the tubing and allow to air dry.
Mask and Headgear
Clean the mask and headgear according to the manufacturer's instructions.
TRoUblesHooTinG
DANGER
Electric Shock Hazard - Do not remove the device cabinet; there are no user-serviceable internal components. The
cabinet should only be removed by a qualified DeVilbiss provider.
If you are having trouble with the operation of your DeVilbiss SleepCube, please refer to the table below for a list of problems with possible causes
and remedies. If you are unable to resolve the problem by following the instructions, contact your home care provider for assistance.
Problem
Probable Cause
Nothing shows on
1. Device is not plugged in or the power cord is
the display.
2. There is no outlet power.
SE-DV54
14
not fully inserted.
troublEShooting/clEaning
Remedy
1a. Verify that power cord is firmly connected to the SleepCube and
the power outlet.
1
If you are using a DC power source make sure the cable
b.
connections are secure. Verify that your battery is charged.
2. Locate a suitable functioning power source.