FujiFilm SonoLite M-Turbo Guia Del Usuario página 16

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 41
Table 6: Manufacturer's Declaration - Electromagnetic Immunity per IEC 60601-1-2:2014
IEC 60601 Test
Immunity Test
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 Vrms in ISM bands
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
Proximity fields
Per 60601-1-2:2014
from wireless
Table 9
communications
equipment
14
Compliance
Electromagnetic Environment
Level
3 Vrms
Portable and mobile RF
communications equipment should be
used no closer to any part of the
6 Vrms in ISM
FUJIFILM SonoSite ultrasound system
bands
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
d
3 V/m
d
80 MHz to
d
2.7 GHz
Where
power rating of the transmitter in
watts (10) according to the
transmitter manufacturer and it is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the vicinity
of equipment marked with the
following symbol:
(IEC 60417 No. 417-IEC-5140:
"Source of non-ionizing radiation")
Per
60601-1-2:2014
Table 9
= 1.2 √
P
= 1.2 √
80 MHz to 800 MHz
P
= 2.3 √
800 MHz to 2.5 GHz
P
is the maximum output
P
b
.
Manufacturer's declaration
a
, should

Publicidad

Tabla de contenido
loading

Tabla de contenido