Calibration gas, factory setting
Zero point
Sensitivity
Calibration gas, optional settings
Calibration gas
Calibration gas concentration
Range limits during calibration of:
Zero point
factory setting
optional settings
Span
factory setting
optional settings
Maintenance signal, factory setting
optional settings
Fault signal, factory setting
optional settings
Beam block warning, factory setting
optional settings
Beam block level, factory setting
optional settings
Warning signal, factory setting
1)
optional settings
Temperature warning, factory setting
optional settings
Response mode, factory setting
optional settings
Compatibility with Dräger REGARD HART
Factory setting
optional settings
4-20 mA output, factory setting
optional settings
Mode 4-20 mA output, factory setting
optional settings
SIL lock, factory setting
optional settings
1) If the warning signal is active, the fault signal is sent for 0.7 seconds every 10 seconds in the event of a warning.
28
Dräger PIR 7000 Type 334
0 %LEL
Methane. 50 %LEL
can be selected within the measured gases
can be selected within the measurement range
33 % (corresponds to 5 % of the configured calibration gas concentration)
0...100 % (corresponds to 0...15 % of the configured calibration gas concentration)
®
50 % (corresponds to 1.5 %LEL)
0...100 % (corresponds to 0...3 %LEL)
constant, 3 mA
constant, 0.7...3.6 mA or alternately, 5 mA for 0.4 s / 3 mA for 0.7 s
<1.2 mA
0,7...3,6 mA
inactive, 2 mA
active/inactive, 0.7 to 3.6 mA
15 %LEL full-scale deflection
0...15 %LEL full-scale deflection
inactive
active/inactive
inactive
active/inactive
normal
normal/fast
disable
on/off
4,0 mA; 20.0 mA
4,0 ± 0,2 mA; 20.0 ± 0.5 mA
autodetect
autodetect / sink / source
disable
on/off
Dräger PIR 7000 Type 340
0 %LEL
Propane, 50 %LEL
can be selected within the
0...100 % (0...0.03 vol. %)
33 % (5 % of the calibration
configured calibration gas conc.)
Dräger PIR 7200
0 % by vol.
Carbon dioxide, 4 vol. %
Carbon dioxide
measurement range
45 % (0.013 vol. %)
gas conc.)
0...100 % (0...15 % of the