Material Resistance - Ivoclar Vivadent UTS 3D Instrucciones De Limpieza Y Esterilizado

Ocultar thumbs Ver también para UTS 3D:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
Validation was carried out on an equivalent product
by an independent accredited and recognized test
laboratory by application of the steam sterilizer HST
6x6x6 (Zirbus technology GmbH, Bad Grund) and the
fractionated vacuum/dynamic air removal procedure
as well the gravity displacement procedure.
The flash/immediate use sterilization procedure must not
be used.
Do not use dry heat sterilization, radiation sterilization,
formaldehyde and ethylene oxide sterilization, as well as
plasma sterilization.
STORAGE
Please store the components after sterilization in the
sterilized packages at a dry and dust-free place.

MATERIAL RESISTANCE

Please take care that the listed substances are not
ingredients of the cleaning or disinfection detergent:
– organic, mineral, and oxidizing acids (minimum admitted
pH-value 6.5)
– lyes (maximum admitted pH-value 8.5, neutral/enzymatic
cleaner recommended)
– organic solvents (for example: acetone, ether, alcohol,
benzine)
– oxidizing agents (for example: peroxide)
– halogens (chlorine, iodine, bromine)
– aromatic, halogenated hydrocarbons
Please do not clean any components using metal brushes or
steel wool.
Please do not use acidic neutralization agents or any rinse
aids.
Please do not expose any components to temperatures
higher than 142 °C (288 °F)!
REUSABILITY
The components can be reused – in case of adequate care
and if they are undamaged and clean – 50 times. The user
is responsible for each further use as well as for the use
of damaged and dirty components (no liability in case of
disregard).
7

Publicidad

Tabla de contenido
loading

Tabla de contenido