Descargar Imprimir esta página

First Time Operation And Initial Settings; Trace Rotation; Probe Adjustment And Use Of Probes; Khz Adjustment - Hameg HM400 Manual

Publicidad

Idiomas disponibles

Idiomas disponibles

If a probe is to be used to contact a BNC connecto, a probe
to BNC adapter should be used in order to prevent groun-
ding problems.
If ripple or noise appears on small signals at high sensitivity settings,
multiple grounds resp. ground loops may cause the problem. The mains
safety earth is quite often the cause, because other test equipment will
also be connected to the same safety earth, this can generate currents
through the shields of cables etc. Most instruments have socalled
Y capacitors connected from the mains to safety earth.

4 First time operation and initial settings

Prior to any use of the instrument make sure the power plug is inserted
before any other contacts are established.
Turn the instrument on by pressing the red button POWER, several
indicators will light up, the oscilloscope will perform a self-test. If any
errors are detected, there will be 5 acoustic signals; it is recommen-
ded to submit the instrument to a service station. After a successful
self-test the instrument will be ready to operate, it will use the settings
stored from its last use.
If there is no trace after a 20 s wait, press the AUTOSET button. If the
trace is now visible, use the SELECT button, the ADJUST + / – buttons to
ADJUST
!
INTENS
set the desired intensity and optimum focus. For best focus adjustment
+
+
FOCUS
it is recommended to display a full screen sine wave. If no input signal
TRACE
is connected, switch the input to GND in order to prevent any noise from
disturbing the focus adjustment.
SAVE
push long
RECALL
POSITION 1
1
1–6
In order to extend the life of the crt, set the intensity no higher than
needed for the specific measurement under the given ambient light
AUTOSET
conditions. Take care not to leave a bright spot as this could burn the
2
crt phosphor. Do not turn the instrument on and off in short intervals.
VOLTS / DIV
It is recommended to always use first the button AUTOSET and select
.2
.1
.5
50
3
the buttons AC and CH 1 in the TRIGGER select area.
VAR
20
1
1
4.1

Trace rotation

10
2
2
In spite of the mumetal crt shield the earth's magnetic field still has
5
5
5
some influence on the crt beam. Depending on the orientation of the
5
2
10
10
instrument the trace may not always remain parallel to the graticule
1
lines. See the chapter Controls for the adjustment of the trace rotation.
20
V
mV
6
DC
DC
4.2
Probe adjustment
and use of probes.
AC
GND
AC
Probes have to be adjusted to the
CH 1
input they are connected to; this
X-INP
adjustment has to be performed
INPUTS
each time a probe is moved to
1 MΩ II 15 pF
!
another input. A generator in
max.
400 V
CAT I
the HM400 delivers a fast rise time (
p
0.2 Vpp the frequency of which can be selected by using the knob TIME/
DIV (see the chapter Controls). The square wave signal is accessible at
the two contacts below the controls. The 0.2 V
probes, sufficient for 4 divisons at 5 mV/DIV.
4.3

1 kHz adjustment

This basic adjustment compensates for the input impedance of the
scope, the probe's capacitor is adjusted so that the capacitive division
equals the resistive division, the division will thus be identical from DC
SELECT
40 MHz ANALOG OSCILLOSCOPE
AUTO
TRIGGER
POSITION 2
LEVEL
NORM
SLOPE
VOLTS / DIV
.2
.1
TRIGGER
.5
50
20
CH 1
10
CH 2
5
2
LINE
1
20
V
mV
EXT
GND
INV
CH 2
PROBE ADJUST
1 kHz / 1 MHz
ca. 0.2 Vpp
<
5 ns) square wave signal of appr.
are destined for 10:1
pp
F i r s t t i m e o p e r a t i o n a n d i n i t i a l s e t t i n g s
to high frequencies. With 1 : 1 probes or probes which can be switched
to 1 : 1, an adjustment is neither necessary nor possible. Before this
adjustment make sure the trace rotation adjustment was performed
(see TRACE rotation).
Connect the 10:1 probe to the input, e.g. CH 1, do not press any button,
set the coupling to DC, the sensitivity with VOLTS/DIV to 5 mV/DIV, the
TIME/DIV switch to 0.2 ms/DIV; make shure that both are calibrated,
i.e. the variables disabled. Connect the probe tip (and ground cable) to
the contact(s) „PROBE ADJUST" (see the photo); a 4 DIV display of two
signal periods should appear. Now adjust the probe capacitor (see the
probe manual for its location) until the square wave is perfectly flat, i.e
there are neither under- nor overshoots. The transitions are invisible at
this sweep speed (see the pictures). The amplitude of the square wave
should be within 4 ±0.12 DIV.
undershoot
4.4

1 MHz adjustment

The probes supplied with the scope have additional adjustment ele-
ments which allow to correct for aberrations at high frequencies.
HM400
After this adjustment maximum bandwidth and best pulse response of
the combination scope and probe are obtained by achieving maximally
flat group delay; overshoots, undershoots, ripple are minimized.
TRIG'd
This adjustment requires a fast square wave generator (
X-POSITION
output impedance (50 ohms) which delivers 0.2 V
ADJUST output of the scope fulfills these requirements.
X-MAG
x10
Connect the 10:1 probe to the input to be used. Select PROBE ADJUST
signal 1 MHz with the knob „TIME/DIV (see the chapter Controls) switch
TIME / DIV
the coupling to DC, the VOLTS/DIV selector to 5 mV/DIV and the TIME/
.2
.1
ms
.5
50
DIV selector to 100 ns/DIV. Connect the probe tip and ground to the two
VAR
20
1
AC
PROBE ADJUST contacts. The square wave will now be visible and also
10
2
its rising and falling slopes. See the probe manual for the location of
5
5
the adjustment elements.
DC
10
2
Adjustment criteria:
20
1
LF
Only the rising slope and the top of the square wave are of concern,
.5
50
disregard the other portions of the signal.
µs
.1s
.2
.2s
.1
TV
Short rise time
HOLD
OFF
Clean transition from the rising slope to the top of the square
ON
Z-INP
wave with no over- or undershoot, flat top.
EXT. TRIG / Z-INP
1 MΩ II
15 pF
max.
100 V
undershoot
p
The amplitude of the square wave should be identical to that with the
1 kHz signal. It is important to always first perform the 1 kHz adjustment,
in general a readjustment of the 1 kHz will not be necessary. Please
note that the probe adjust frequencies are not precise and hence must
not be used for any checks of the accuracy of the time base, also their
duty cycle is not controlled.
The probe adjust signal must conform to the requirements of zero
potential at the bottom of the square wave, precise amplitude and flat
tops; its frequencies and duty cycles need not be precise.
correct
overshoot
<
5 ns) and low
at 1 MHz; the PROBE
pp
correct
overshoot
Subject to change without notice
33

Publicidad

loading