Jeopardising the Therapy Goal Due to Lack of Free Movement
Check if the system joint moves freely in order to avoid restrictions of the joint function. Use suitable
sliding washers according to the information in this manual.
Limitation of the Joint Function Due to Improper Processing
Errors in processing can impair the joint function. Pay particular attention to:
- correctly connecting the system side bar/system anchor with the system case in accordance with the
production technique;
- greasing the joint components only slightly and
- adhering to the maintenance intervals.
Damages to Batteries Due to Improper Handling
Use the batteries as described in this manual. Pay particular attention to:
- using the batteries with the delivered charger;
- only using the batteries at temperatures of -20°C to +45°C;
- only using undamaged batteries (without leakage of substances/liquids) and
- properly disposing of the batteries.
3. Application
The NEURO TRONIC knee joint system with component set including system knee joint and controller is
exclusively for use for orthotic fittings of the lower extremity. It must be handled by a professionally trained
user. All FIOR & GENTZ system joints were developed for everyday life activities such as standing and walking.
Extreme loads connected to activities like running, climbing and parachuting are excluded.