7. Setting-up the System

7.1. EQUIPMENT VISUAL CHECK

Make sure that the Monitor, accessories and cables are clean
and in correct mechanical condition.
WARNING
Do not use the Pressio
tect any defect, such as cracks, frayed cables or broken
connectors, on the Monitor or its cables or accessories.

7.2. POSITIONING THE SYSTEM

Position the Monitor horizontally, using the pole clamp to
prevent any forward tilting. If the Monitor must be placed on
a surface, make sure the surface is flat.

8. Cleaning and disinfection procedures

CAUTION
Do not use the Monitor, its cables or accessories if
there is any soiling or visible residues on the external
surface of the components or inside the cable wires.
1. Power off the Monitor. If the Monitor is operating on
mains power, unplug it as well.
2. Disconnect all the cables and USB devices from the
Monitor.
NOTICE
Avoid any contact with the connectors on the right pan-
el of the Monitor.

8.1. INTRODUCTION

The Monitor and the cables are delivered clean, but not dis-
infected.
Clean the Monitor and all its cables before first use and be-
tween each patient, as described here. Then disinfect the
Catheter Extension Cable.
Page 18 of 174
button.
®
Monitoring System if you de-
The Monitor is now ready to use.
NOTE
For maximum battery life, first operate the Monitor on
mains power for several hours, until battery charge indi-
cator light appears near the right end of the status bar,
at top right corner of the Monitor touch screen:
CAUTION
Use the pole clamp to immobilize the Monitor during
monitoring and patient transport, to prevent an acci-
dental fall of the Monitor, and possible traction on the
catheter extension cable and the implanted Catheter.
WARNING
Do not clean the Monitor or the Catheter, nor disinfect
the cables, when in use on the patient.
WARNING
Do not immerse, autoclave or soak the Monitor or its
accessories in a liquid. Their performance (including
drift and electrical safety) could be affected.
WARNING
Catheter kits are supplied sterile, for single use only.
Do not reuse a Catheter.
WARNING
Do not re-sterilize a Catheter, or re-use it after opening
the packaging, and/or after explantation.
For more information, refer to the Instructions For Use
(IFU) of the Catheter being used.