Guidance and manufacturer's declaration – Electromagnetic im-
munity
Proximity
Band/Test level
fields
from
380 - 390 MHz /
RF wireless
27 V/m
communica-
430 - 470 MHz /
tions equip-
28 V/m
ment
704 - 787 MHz /
IEC
9 V/m
61000-4-3
800 - 960 MHz /
28 V/m
1,700 - 1,990
MHz / 28 V/m
2,400 - 2,570
MHz / 28 V/m
5,100 - 5,800
MHz / 9 V/m
RATED
30 A/m
power
fre-
50 Hz or 60 Hz
quency
magnetic
fields
IEC
61000-4-8
Electrical
±2 kV
fast
transi-
100 kHz repeti-
ents / bursts
tion frequency
IEC
61000-4-4
Surges line-
±0.5 kV, ±1 kV
to-line
IEC
61000-4-5

19. References

®
Table 1. Pressio
Catheter Kits
®
PSO-PB
Pressio
ICP monitoring kit, parenchymal with bolt
®
PSO-PBT Pressio
ICP and ICT monitoring kit, parenchymal with
bolt
®
PSO-PT
Pressio
ICP monitoring kit, parenchymal tunneling
®
PSO-PTT Pressio
ICP and ICT monitoring kit, parenchymal tunnel-
ing
®
PSO-VT
Pressio
ICP monitoring kit, ventricular tunneling
with external CSF drainage function
®
PSO-VTT Pressio
ICP and ICT monitoring kit, ventricular tunneling
with external CSF drainage function
®
Table 2. Pressio
Monitoring System
®
PSO-4000 Pressio
2 ICP Monitor
Power cable and catheter extension cable (PSO-EC30) in-
cluded
Page 34 of 174
cluding
Band/Test level
bles specified
380 - 390 MHz /
by the manu-
27 V/m
facturer. Oth-
430 - 470 MHz /
erwise, deg-
28 V/m
radation
704 - 787 MHz /
the perform-
9 V/m
ance of this
800 - 960 MHz /
equipment
28 V/m
could result.
1,700 - 1,990
Interference
MHz / 28 V/m
occur in the vicinity
2,400 - 2,570
of equipment con-
MHz / 28 V/m
taining a RF trans-
5,100 - 5,800
mitter.
MHz / 9 V/m
30 A/m
Power
frequency
magnetic
50 Hz or 60 Hz
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital environ-
ment.
±2 kV
The quality of the
electricity
100 kHz repeti-
network should be
tion frequency
that of a typical
commercial or hos-
pital environment.
±0.5 kV, ±1 kV
The quality of the
electricity
network should be
that of a typical
commercial or hos-
pital environment.
Guidance and manufacturer's declaration – Electromagnetic im-
munity
ca-
Surges line-
to-ground
IEC
61000-4-5
of
Conducted
disturban-
ces induced
by RF fields
may
IEC
61000-4-6
Voltage dips
fields
IEC
61000-4-11
supply
Voltage
in-
terruptions
IEC
supply
61000-4-11
NOTE
U
is the AC mains voltage prior to application of the test level.
T
Table 3. Pressio
PSO-EC30
PSO-MCxx
PSO-MCT-y
PSO-DR
PSO-MRI
±0.5 kV, ±1 kV,
±0.5 kV, ±1 kV,
±2 kV
±2 kV
3 V
3 V
0.15 MHz – 80
0.15 MHz – 80
MHz
MHz
6 V in ISM
6 V in ISM
bands between
bands between
0.15 MHz and
0.15 MHz and
80 MHz
80 MHz
80 % MA at
80 % MA at
1 kHz
1 kHz
0 % U
; 0.5 cy-
0 % U
; 0.5 cy-
T
T
cle
cle
At 0°, 45°, 90°,
At 0°, 45°, 90°,
135°, 180°,
135°, 180°,
225°, 270° and
225°, 270° and
315°
315°
0 % U
; 1 cycle
0 % U
; 1 cycle
T
T
and
and
70 % U
; 25/30
70 % U
; 25/30
T
T
cycles
cycles
Single phase: at
Single phase: at
0 % U
;
0 % U
;
T
T
250/300 cycles
250/300 cycles
®
Accessories
Catheter extension cable (included with PSO-4000, al-
so available separately)
Pressure cable
– MC01: PHILIPS (AGILENT) - 12 pins
– MC02: SIEMENS (SIRECUST) - 10 pins
– MC03: SPACELABS & MINDRAY - 6 pins
– MC04: GE DATEX-Ohmeda - 10 pins
– MC05: GE Solar (MARQUETTE) - 11 pins
– MC06: HELLIGE - 10 pins
– MC07: SIEMENS - 7 pins
– MC08: NIHON KOHDEN - 5 pins
– MC10: DATASCOPE - 6 pins
Temperature cable
– MCT-A: PHILIPS (AGILENT) - 2 pins
– MCT-B: SIEMENS - 7 pins
– MCT-C: SPACELABS - 10 pins
– MCT-E: GE Solar (MARQUETTE), GE DATEX-Ohme-
da - 11 pins
– MCT-F: HELLIGE, DATEX-Ohmeda, NIHON KOH-
DEN, MINDRAY & DATASCOPE - JACK 6.35 mm
Single use, disposable drill
®
Pressio
MRI Support
The quality of the
electricity
supply
network should be
that of a typical
commercial or hos-
pital environment.
It is possible that
ripples may appear
on the ICP curve.
Keep the source of
disturbance
away
from the cables of
the Monitor.
The quality of the
electricity
supply
network should be
that of a typical
commercial or hos-
pital environment.
If the user of the
Monitor
requires
continuous opera-
tion during power
cuts, it is recom-
mended that the
Monitor is supplied
by
an
energy
source not subject
to cuts, or supplied
by a battery.