SYMPTOMS
POSSIBLE CAUSES SUGGESTED ACTIONS
The flashing mes-
The catheter exten-
sage "CONNECT
sion cable connector
CATHETER" per-
is not completely
sists after connect-
pushed into the
ing the Monitor with
socket on the Moni-
the catheter exten-
tor.
sion cable and the
The Catheter is not
Catheter.
correctly connected
to the catheter ex-
tension cable.
Catheter extension
cable damaged.
Catheter damaged.
Loss of monitoring
with connected
Electrostatic dis-
catheter, or Monitor
charges
malfunction

17. Technical Specifications

17.1. GENERAL POINTS

The catheter extension cable and the Catheters are the ap-
plied parts of the Monitor.
Monitor general specifications
Item
Specification
Type of monitor
Continuous ICP/ICT monitor
Dimensions
L198 mm x H127 mm x D106 mm
Weight
1.8 kg (4 lbs)
Voltage
Power supply
Consumption
Frequency
Power cable
Length
Type
Voltage
Capacity
Energy
Internal battery
Weight
Safety
Dimension
Page 28 of 174
Ensure the catheter ex-
tension cable is correctly
connected to the Moni-
tor.
Check the catheter ex-
tension cable is correctly
connected to the Cathe-
ter.
Replace the catheter ex-
tension cable.
Replace the Catheter.
Power the Monitor off
and on.
If the Power button does
not answer, unplug the
power cable, remove the
battery cover and dis-
connect the battery for a
few seconds.
Reconnect the battery,
put the battery cover
back, plug the power ca-
ble back and power on
the Monitor.
If the Monitor still freezes
on the Home page, re-
turn the Monitor to after-
sales service.
100-240 V~
12-80 VA
50-60 Hz
– Europe, Switzerland, Brazil, Chi-
na, South Africa, Australia: 2.50 m
– Japan: 2.30 m
– USA: 3 m
Lithium ion
3.65 V
8.0 Ah
29.2 Wh
210 g (0.5 lbs)
IEC 62133:2017; UL 1642; UN/DOT
38.3
L92 mm x H70 mm x D19 mm
SYMPTOMS
POSSIBLE CAUSES SUGGESTED ACTIONS
Monitor general specifications
Item
Specification
Case
Material
Cooling
By convection
Pressure sen-
sor
Measurement
method
Temperature
sensor
5.7" VGA colour LCD TFT
Display
Resolution
Viewing angle 80° min
Type
Touch screen
Material
Accessories and compatible catheters length
Item
Accessories
Catheter Extension Cable (PSO-EC30)
Pressure cable (PSO-MCxx)
Temperature cable (PSO-MCT-y)
Mini-USB cable
If error codes "E001",
"E002" or "E005" ap-
pear:
Disconnect and recon-
nect the catheter.
If needed, restart the
Monitor.
If needed, change the
catheter.
®
BASF Ultradur
FRee B 4450 G5 LS
25% Glass Filled PBT
Strain gauge piezoresistive sensor.
Differential sensor (atmospheric ref-
erence)
YSI400 Thermistor
640 x 480 pixels
Capacitive
Glass
Length
2 m
2.90 m
2.90 m
1.80 m