17.2.1. Physiological alarms

Physio-
Clinical
Response
logical
Priority
time
parame-
ter
ICP
High
Prompt
ICT
Medium
Delayed

17.2.2. Technical alarms

Techni-
Alarm
Priori-
Re-
cal pa-
condi-
ty
sponse
rameter
tion
time
Pressure
Fault in
High
Prompt
sensor
the
status
Catheter
pressure
sensor
Sound indica-
Visual indica-
tors
tors
Three-pulse
The ICP display
sound, pause
area flashes red
for 300 ms,
at a frequency
two-pulse
of 2 Hz.
sound, pause
The ICP curve
for 500 ms,
cursor turns
three-pulse
red.
sound, pause
An arrow indi-
for 300 ms,
cates the direc-
two-pulse
tion in which
sound
the threshold is
The pulses are
exceeded on
separated by
the ICP alarm
100 ms
scale.
Sequence repe-
The word
tition interval:
ALARM dis-
2.5 s
plays in the ICP
The pulse vol-
display area.
ume is 55 dB
Three-pulse
The ICT display
sound
area flashes
yellow at a fre-
The pulses are
quency of 0.5
separated by
Hz.
100 ms
An arrow indi-
Sequence repe-
cates the direc-
tition interval:
tion in which
7.5 s
the threshold is
The pulse vol-
exceeded on
ume is 55 dB
the ICT alarm
scale.
The word
ALARM dis-
plays in the ICT
display area.
Sound indica-
Visual indi-
tors
cators
Three-pulse
The mes-
sound, pause
sage E001
for 300 ms,
ALARM
two-pulse
displays.
sound, pause
The display
for 500 ms,
area flash-
three-pulse
es red at a
sound, pause
frequency
for 300 ms,
of 2 Hz.
two-pulse
sound
The pulses are
separated by
100 ms
Sequence rep-
etition interval:
2.5 s
The pulse vol-
ume is 55 dB
Techni-
Alarm
Priori-
Re-
cal pa-
condi-
ty
sponse
rameter
tion
time
Tempera-
Fault in
High
Prompt
ture sen-
the
sor sta-
Catheter
tus
temper-
ature
sensor
Calibra-
Catheter
High
Prompt
tion sta-
calibra-
tus
tion fault
ICP
Average
High
Prompt
measure-
ICP out-
ment
side
range
meas-
urement
range
Sound indica-
Visual indi-
tors
cators
Three-pulse
The mes-
sound, pause
sage E005
for 300 ms,
ALARM
two-pulse
displays.
sound, pause
The display
for 500 ms,
area flash-
three-pulse
es red at a
sound, pause
frequency
for 300 ms,
of 2 Hz.
two-pulse
sound
The pulses are
separated by
100 ms
Sequence rep-
etition interval:
2.5 s
The pulse vol-
ume is 55 dB
Three-pulse
The mes-
sound, pause
sage E002
for 300 ms,
ALARM
two-pulse
displays.
sound, pause
The display
for 500 ms,
area flash-
three-pulse
es red at a
sound, pause
frequency
for 300 ms,
of 2 Hz.
two-pulse
sound
The pulses are
separated by
100 ms
Sequence rep-
etition interval:
2.5 s
The pulse vol-
ume is 55 dB
Three-pulse
The mes-
sound, pause
sage E001
for 300 ms,
ALARM
two-pulse
displays.
sound, pause
The display
for 500 ms,
area flash-
three-pulse
es red at a
sound, pause
frequency
for 300 ms,
of 2 Hz.
two-pulse
sound
The pulses are
separated by
100 ms
Sequence rep-
etition interval:
2.5 s
The pulse vol-
ume is 55 dB
Page 31 of 174