6.4 - Brushless Electrical Micromotor - Side Panel (Optional)
Before starting the operation, connect the
handpiece to the electrical micromotor (16).
To
activate
propulsion/activation pedal (14) or the Manual Start
key (11).
To select a preset program: use the Program
key (8). There are 7 programs available, according to
the table below:
The selected program is shown on the display (1).
To record program: select program P1 or P2, then select the desired
gear ration and adjust the speed. Press and hold the Program key (8) until
the confirmation of the system.
To adjust speed: use the adjustment keys (13) to increase or decrease speed. Please see the table below
for the speed range for each gear ration value.
The selected program is shown on the display (5).
To adjust the micromotor water: use the water control knob (15)
on the instruments' control panel.
To access the setup menu: press and hold the Setting key (12).
Use the Setting key (12) to select the desired option.
To adjust LED options: use the adjustment keys (13) to set the
parameter.
LED Off
To adjust the pneumatic activation method: use the adjustment keys (13) to set the parameter.
Micromotor performs speed according to the
pressure on the pedal (14)
To adjust Input pneumatic pressure: use the adjustment keys (13) to set the parameter.
Set minimum activation pressure
from 0.03 to 0.10 MPa
Note: the micromotor system preset by the factory according to the internal equipment pressure. It is not
recommended to change the pressure parameters.
the
rotation:
press
To select the gear ratio: use the Gear Ratio key (10). There are the
following values available: 1:5, 1:1, 4:1, 10:1, 16:1 and 20:1.
The selected gear ratio is shown on the display (4).
To change the rotation direction: use the Direction key (9). The
selected direction is shown on the display (3).
LED On
LED On
without delay
with 10s delay
When pedal (14) is pressed micromotor
performs the maximum speed adjusted
the
Reverse
Set maximum input pressure
from 0.20 to 0.40 MPa
17
Forward