Parameter
Trigger
Cycle
Max IPAP
Min EPAP
Pressure
Support
Easy-
Breathe
(US Only)
Max
Pressure
Min
Pressure
Min EPAP
Max EPAP
Options
Climate
9 9 9 9 9 9 9 9 9 9 9 Auto / Manual
Control
Therapy LED 9 9 9 9 9 9 9 9 9 9 9 On / Off
18
Modes
9 9 9
9 9
9 9 9
9
9
9
9
9 9
9 9
Range
Very Low / Low /
Med / High / Very
High
Default: Med
Very Low / Low /
9
Med / High / Very
High
Default: Med
4–25 cm H
2
Default:
25 cm H
O
2
4–[Max IPAP]
cm H
O
2
Default:
4 cm H
O
2
0–10 cm H
2
Default:
4 cm H
O
2
On / Off
Default: On
Min–20 cm H
0.2 cm H
O
2
increments
Default:
20 cm H
O
2
4–Max cm H
0.2 cm H
O
2
increments
Default:
4 cm H
O
2
9 4–15 cm H
2
Default:
4 cm H
O
2
9 4–15 cm H
2
Default:
15 cm H
O
2
Default: Auto
Default: Off
Description
Sets the level of inspiratory
flow above which the device
changes from EPAP to IPAP .
Sets the level of inspiratory
flow below which the device
changes from IPAP to EPAP .
O
Sets the maximum
inspiratory pressure
delivered by the device.
Sets the minimum expiratory
pressure delivered by the
device.
O
Difference between IPAP
and EPAP . Adjust for patient
comfort.
Enables Easy-Breathe
comfort feature.
O,
Sets the upper limit of
2
treatment pressure.
O,
Sets the lower limit of
2
treatment pressure.
O
Sets the minimum EPAP
delivered by the device.
O
Sets the maximum EPAP
delivered by the device.
Dependent on Min EPAP .
Sets the type of Climate
Control.
Enables or disables the blue
LED.