Using the Screening Report to Determine an
Acceptable Sense Vector

Report Description

The EMBLEM S-ICD Automated Screening Report PDF summarizes
screening results and instructs users to check QRS complex morphology
across postures to identify acceptable leads. The ECG data used by the
AST is provided for each lead/posture combination tested.
NOTE: The gain of the ECG signal for each vector is automatically
adjusted prior to printing. This allows the printed report ECGs to be
assessed with the Model 4744 Patient Screening Tool, if desired.

Saving, printing and exporting the report

1. To save the report, click the Save Report button on the lower
left-hand corner of the Screening tab - a PDF of the report is
saved to the Model 3120 Programmer hard drive.
NOTE: It may take a couple of minutes to generate the PDF
report, depending on the number of postures or conditions run.
NOTE: A maximum of 100 reports may be saved to the hard
drive. A capacity warning appears when the maximum has been
reached. The user has the option to proceed with the oldest
session being deleted or abort the session.
2. To print the report, click the Print Report button to send the
report to a connected external printer.
3. To export the data to a pen drive:
▪ Click End Screening Session to return to the programmer
home screen.
▪ Insert pen drive into the USB port of the Model 3120
programmer.
▪ Click Patient Data Management, then select the Export
Screening Data tab.
▪ Click Export or Export with Password Protect and follow the
prompts (all saved reports are exported).