Selecting the Distraction System
WARNING
Danger of injury due to failure of the distraction system!
Gebrüder Martin distraction systems must not be combined with systems from other
manufacturers!
Using instruments not specifically intended for use with implants can result in failure of the
distraction system!
Distractor
Illustration
Patient screwdriver and flat spanner
Illustration
32
Description
Starting position: 9 mm (distance from plate to plate)
Maximum distraction distance: 19 mm
Application in case of extreme maxilla crowdings
Starting position: 13 mm (distance from plate to plate)
Maximum distraction distance: 19 mm
Application in case of regular maxilla crowdings
Description
Patient screwdriver, straight
Patient screwdriver, angled
Activation key, double-bent
Flat spanner for counter nut
Instructions for Use
Rotterdam Palatal Distractor
Item no.
51-555-09-09
51-555-13-09
Item no.
51-555-85-07
51-555-95-07
51-555-90-07
51-555-91-07