Change of batteries
Batteries are consumables and must be exchanged regularly.
For this purpose, remove the cover of the battery case at the
rear of the unit and exchange the batteries. Subsequently,
remount the cover of the battery case. Dispose of unserviceable
batteries according to the national rules and regulations.
Maintenance and cleaning
Clean the radiometer with a soft cloth and customary aldehyde-free
disinfecting solution. Do not clean with highly aggressive disinfecting
solutions (e.g. solutions based on orange oil or with an ethanol
content of more than 50%), solvents (e.g. acetone), or pointed
instruments, which may damage or scratch the plastic.
– Do not spray disinfecting solution directly on the apparatus and
do not hold it under running water in order to prevent liquids
from entering the device.
– Do not sterilize the radiometer in the autoclave.
Warranty
The warranty for the bluephase meter is three years from the date of
purchase. Malfunctions resulting from faulty material or
manufacturing errors are repaired free of charge during the warranty
period. The warranty does not provide the right to recover any
material or non-material damage other than the ones mentioned. The
apparatus must only be used for the intended purposes. Any other
uses are contraindicated. The manufacturer does not accept any
liability resulting from misuse and warranty claims cannot be
accepted in such cases. This is particularly valid for:
– damage resulting from improper handling.
– damage to components resulting from wear under standard
operating conditions (e.g. sensor).
– damage resulting from external influences, e.g. blows, drop to the
floor;
– damage resulting from improper repairs or modifications that
have not been carried out by certified Service Centers.
In case of a claim under warranty, the apparatus must be returned,
carriage paid, to the dealer or directly to Ivoclar Vivadent in its
original packaging together with the purchase document.
Procedure in case of repair
Repair work may only be carried out by a certified Ivoclar Vivadent
Service Center. In case of a defect that cannot be rectified, please
contact your dealer or your local Service Center (see addresses on the
reverse side). A clear description of the defect or the conditions under
which the defect occurred will facilitate locating the problem. Please
enclose this description when returning the apparatus.
Product specifications
Delivery form
– 607922 bluephase meter
– 611504 Operating Instruction
– 3x Batteries (Type LR6/AA/1.5V)
Technical data
Operating voltage
Battery operation
Measuring range:
Wavelength
Power density
Only for light probes with a circular emission window
Measuring accuracy
Operating conditions:
Temperature
Relative humidity
Ambient pressure
Dimensions
Weight
Transportation and storage conditions
– Temperature -20 °C to +70 °C
– Relative humidity 10% to 75%
– Ambient pressure 500 hPa to 1060 hPa
– bluephase radiometer has to be stored in closed, roofed rooms.
Protect the device from severe jarring.
4.5 VDC
3x Type LR6/AA/1.5V
380–515 nm
300–2500 mW/cm
Ø 7-13 mm
+/- 20%
+10 °C to +40 °C
30% to 75%
700 hPa to 1060 hPa
L = 126 mm W = 80 mm H = 41 mm
2
170 g