Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The A-71100-XX/ A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A‑ 71101-XX should make
sure it is used in such an environment.
Immunity Test
Power input line
voltage drop, instant
supply interruption
and voltage
fluctuation
IEC 61000-4-11
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
58
IEC 60601
Test Level
< 5% U
T
(> 95% dip in U
)
T
for 0.5 cycles
40% U
T
(60% dip in
U
) for 5 cycless
T
70% U
T
(30% dip in U
) for
T
25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 seconds
3 A/m
Allen® Advance Table User Manual (D-720761 Ver D)
Electromagnetic
Compliance
Environment—
Level
Guidance
< 5% U
Mains power quality
T
(> 95% dip in U
)
should be of a typical
T
for 0.5 cycle
commercial or hospital
40% U
environment. If the user
T
(60% dip in U
) for
of the A-71100-XX/
T
5 cycles
A-71101-XX requires
70% U
continued operation
T
(30% dip in U
) for
during power mains
T
25 cycles
interruption, it is
< 5% U
recommended that the
T
(> 95% dip in U
)
A-71100-XX/
T
for 5 seconds
A-71101-XX be
powered from an
uninterruptedly power
supply or a battery.
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.