Stryker Formula Manual Del Usuario página 15

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 191
3. If applicable, press the "OK" button on the shaver console warning screen.
The default home screen will appear.
Installing and Removing Shaver Blade
Warning
Formula® Handpieces
1. Align the locking guides on the base of the shaver blade with the locking
guides on the handpiece, as shown below.
2. Slide the blade into the handpiece until it locks into place.
3. Gently tug the blade to ensure that it is engaged.
4. To remove a blade, press the quick-release button on the handpiece (A)
and withdraw the blade (B), as shown below.
• Do not bend shaver blades or use them for prying.
Excessive pressure on them may cause it to break and
may cause harm to the patient and operating-room staff.
Broken pieces may also enter the surgical site, where they
will be difficult to remove.
• Avoid contacting shaver blades with metal objects. Do
not insert metal objects in the shaver blade window. If
shaver blades contact metal objects, they may break,
leaving pieces in the surgical site where they will be
difficult to remove.
B
A
EN-11

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido