Automated
Cleaning
12
Cleaning was
validated using
Enzol® at 1 oz/gal. at
35-40°C.
13
Cleaning was
validated using
Neodisher Mediclean
Forte
at 5 ml/L at
TM
30°C.
9. Inspect
•
Visually inspect the handpiece, including all
internal surfaces, for remaining soil. Use an
endoscopic camera and endoscope if necessary
to see the inner surface of the lumen.
•
If soil remains, repeat the manual cleaning
procedure, focusing on those areas.
1. Wipe
Wipe excess soil from the handpiece with disposable
paper towels.
2. Soak in Detergent
Note: In the U.S., use an enzymatic detergent. In the
EU, use Neodisher Mediclean Forte.
U.S. only: Prepare an enzymatic detergent
•
solution
manufacturer's recommendations.
•
EU only: Prepare detergent solution
lukewarm tap water, according to manufacturer's
recommendations.
•
Wipe the entire surface of the handpiece using a
soft clean cloth dipped in the detergent solution.
•
Immerse the handpiece in the detergent solution,
ensuring the solution contacts all inner and outer
surfaces of the handpiece. Use a syringe to inject
lumens with 50mL of solution.
•
Soak the handpiece in the solution for 15 minutes.
with lukewarm tap water, according to
12
with
13
EN-23