Cleaning
1
Cleaning was
validated using
Enzol® at 1 oz/gal. at
35-40°C.
2
Cleaning was
validated using
Neodisher Mediclean
Forte
at 5 ml/L at
TM
30°C.
1. Wipe
Wipe excess soil from the handpiece with disposable
paper towels.
2. Soak in Detergent
Note: In the U.S., use an enzymatic detergent. In the
EU, use Neodisher Mediclean Forte.
U.S. only: Prepare an enzymatic detergent
solution
manufacturer's recommendations.
EU only: Prepare detergent solution
lukewarm tap water, according to manufacturer's
recommendations.
Wipe the entire surface of the handpiece using a
soft clean cloth dipped in the detergent solution.
Immerse the handpiece in the detergent solution,
ensuring the solution contacts all inner and outer
surfaces of the handpiece. Use a syringe to inject
lumens with 50mL of solution.
Soak the handpiece in the solution for 15 minutes.