Periodic Maintenance Schedule
To ensure the safety and longevity of your handpiece, observe the following
maintenance procedures:
Caution
To ensure proper heat dissipation and keep the joints clear, do not run the
handpiece without proper suction or irrigation. Immerse the blade in a
sterile irrigant and activate suction while testing the handpiece.
Before Every Surgical Procedure
1. Prior to inserting shaver blade, run the handpiece through its various
functions, modes, and settings to ensure it responds properly to
commands. Ensure that all parts move freely and that there are no loose
or missing components. Note any unusual sounds or vibrations, and note
operating speeds.
2. Refer any calibration or operating difficulties to your Stryker
representative, distributor, Stryker customer service, or the nearest Stryker
subsidiary.
Every 12 Months
1. Check the leakage current using a true RMS digital multimeter and safety
analyzer. Consult the analyzer's instructions for details.
2. When storing or shipping the handpiece, package the handpiece in its
original packaging materials.