Stryker Formula Manual Del Usuario página 26

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 191
8
Cleaning was
validated using a
Spectrum M16 brush.
9
Cleaning was
validated using a
10mm x 24" sized
lumen brush.
10
Cleaning was
validated using
reverse osmosis
deionized (RO/DI)
water.
11
Drying was
validated with
pressurized air at
20 psi.
EN-22
6. Brush
Thoroughly brush the exterior of the handpiece,
including mated and rough surfaces, with a soft-
bristled brush.
Brush around all movable parts, including the
suction valve, while moving them through all
extreme positions, as shown below.
Inject all lumens, crevices, and mated surfaces five
times with 50mL of detergent solution.
With the handpiece immersed in the detergent,
thoroughly brush the interior and exterior with an
appropriately sized soft-bristled brush.
appropriately sized brush for the interior of the
handpiece will be a bottle brush with bristles that
extend the diameter of the handpiece lumen.)
Brush the handpiece lumens while actuating the
suction lever through its extreme positions.
7. Rinse
Remove the handpiece from the detergent
solution and rinse it with water
temperature until all visible detergent residue is
removed.
Flush all lumens, crevices, and mated surfaces five
times with 50mL of water.
After all detergent residues are removed, rinse the
handpiece for an additional 30 seconds.
Drain any excess water from the handpiece by
holding it at an incline.
8. Dry
Dry the handpiece using a clean cloth. Filtered
pressurized air
8
can be used to assist drying.
11
(Note: An
9
at ambient
10

Publicidad