8.1. Specifications

Ultrasound
Frequency
Read out
Contact control level
Timer
Ultrasound continuos
Pulse frequency / duty cycle
Intensity
Ultrasound pulsed mode
Pulse frequency / duty cycle
Intensity
Treatment heads
1 and 3 MHz, large transducer (standard)
1 and 3 MHz, small transducer (optional)
* SATP
= Spatial Average Temporal Peak (average pulsed power)
* ERA
= Effective Radiating Area, this is the effective radiating area of the treatment head.
* BNR
= Beam Nonuniformity Ratio, indicates the ratio between the peaks and the average value of the
intensity in the sound beam. A low BNR excludes high, unwanted energy concentrations.
ERA and BNR are measured confirm the FDA- method. All treatment heads have a visual contact control.
Electrotherapy general
Channels
Current characteristic
Setting resolution
Timer
Polarity reversion direct currents
The maximum intensity within the specification is achieved up to a load of 500 Ω (CC).
Symmetrical and asymmetrical biphasic Pulsed Current, TENS
Frequency
Phase duration (pulse width)
Burst frequency
Modulation frequency (spectrum)
Modulation time (spectrum)
Surge pattern
Intensity
In certain pulse width, frequency and spectrum combinations the setting range is reduced to 90-mA max. in
order to meet the limit values of IEC 60601-2-10

CHAPTER 8 - SPECIFICATIONS

:
1 and 3 MHz
:
intensity in W/cm² and power in W (SATP*)
:
65%
:
0 - 30 minutes, linked to the contact control
:
100 Hz / 100 %
:
0 - 2 W/cm²
:
100 Hz / 5, 10, 20, 50, 80 %
:
0 - 3 W/cm², duty cycle 5, 10, 20, 50 %
:
0 - 2.5 W/cm², duty cycle 80 %
:
Geometric surface 5.8 cm², ERA* 5 cm², BNR* max.
5, type collimating. Side panel radiation max. 10 mW/
cm² (IEC requirement < 100 mW/cm²)
:
Geometric surface 1.15 cm², ERA* 0.8 cm², BNR* max.
5, type collimating. Side panel radiation max. 10 mW/
cm² (IEC requirement < 100 mW/cm²)
:
2
:
Constant Current (CC) or Constant Voltage (CV)
:
0.2 mA
:
0 - 60 minutes
:
manual
:
1 - 200 Hz
:
10 - 400 µs
:
0 - 9 Hz
:
0 - 180 Hz
:
1/1, 6/6, 12/12 en 1/30/1/30 s
:
Ramp up 0 - 9 s
Hold time
1 - 60 s
Ramp down
0 - 9 s
Interval time
1 - 60 s
Delay time
0.1 - 80 s
:
0 - 140 mA.
47