Programming The Pump - Nutricia Flocare Infinity Instrucciones De Uso

Ocultar thumbs Ver también para Infinity:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 31
UK-IE
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols, words and a back light.
The following information can be found in the display:
• Flow rate (ml/h), volume (ml), is displayed through the large characters.
Words below describe what the number relates to (rate, dose or volume).
The pump also displays messages, for example "end of dose" will appear when the pump has finished
delivering a single feed dose.
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
• The Flocare
Infinity
®
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see paragraph: filling the pump
set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "CLR" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
• We strongly recommend that for those patients where interruptions or delay in therapy could affect their health status, the
DOSE setting of the Infinity pump must be used and programmed.
ON/OFF
d
ml/h
• Adjust the flow rate (ml/h) if required by using the "+" or "-" key. Hold either key down to change rapidly.
• If another parameter is indicated in the LCD, it will be necessary to first press the "ml/h" key followed by the "+" or "-" key to set
the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortly at 50 - 125 - 250 ml/h.
ZMNL900911_AW.indd 77
ZMNL900911_AW.indd 77

PROGRAMMING THE PUMP

pump can be used for continuous or dose feeding.
CONTINUOUS FEEDING
+
400 ml/h
d
d
START/STOP
-
1 ml/h
DISPLAY
• The arcs around the run symbol rotate when the pump is running.
• The remaining battery capacity (see paragraph: battery
operation).
• The wall plug symbol indicates that the charger is plugged in.
-77-
3.06.21 9:01
3.06.21 9:01

Publicidad

Tabla de contenido
loading

Tabla de contenido