• When another feeding program needs to be started, or the pump will be used for another patient, all settings (rate and dose)
can be returned to default and the volume delivered can be set at zero by simply pressing the "CLR" key and keeping it
pressed for 2 seconds.
• The Flocare
Infinity
®
™
Performance may degrade due to age, excessive temperatures and number of charging cycles.
• The battery symbol in the display is automatically replaced by the plug symbol when the charger is plugged in. The "fuel
gauge" then indicates that the battery is charging by displaying the bars in an ascending low to high pattern starting with
the leftmost bar.
This pattern continuously repeats while the pump is charging.
• To check the status of the battery, disconnect the charger from the pump and turn the pump on. The bars
between E and F (E=Empty, F=Full) represent the "fuel gauge" of the battery. Each bar is approximately 1/4
of a full battery charge. If 2 bars appear the battery is half full and remaining operating time is approxima-
tely 12 hours at a flow rate of 125 ml/h for a new battery.
• In case of power failure, when the pump is connected to mains, the pump automatically switches to battery powered ope-
ration.
• When during functioning the last bar is gone, the battery symbol will blink to indicate that there is approximately 1 hour of charge
left. The display will flash "BATT" every 3 seconds, alternating with the active display and the pump will beep every 2 seconds to
remind the user of the low battery charge condition. Plug in the charger to continue to run and recharge the battery.
• In case of battery failure, caused by, for example, an excessive temperature condition (which may resolve itself with time) or by a
complete battery failure, the E, F and battery symbol will flash. Turn the pump "off" and contact your healthcare provider or refer
the pump to service.
The power connector at the side of the pump offers the option of connecting the pump to an external alarm system or Patient
Data Monitoring System (PDMS).
• The Flocare
Infinity
®
™
alarm system may be with open or closed contact. Follow the instructions for use of the Flocare
nect the pump.
• The Flocare
Infinity
®
™
instructions for use of the Flocare
• For good pump operation, the pump should be kept clean and free of debris like (dried) tube nutrition. It is recommended to clean
the pump with a wet cloth after every finished feeding regimen and specifically in between patient uses (if applicable).
• Detergent and disinfectant cleaning agents and wipes can damage plastic surfaces of medical devices when they are not com-
patible with the surface material. Damaged surfaces may compromise the ability to decontaminate medical devices adequately
and / or may interfere with device function and affect their integrity and performance. Ensure cleaning solutions, detergents and
disinfectant wipes are compatible with the device. Look for signs of damage, cracking, colour changes, etc. to the medical
ZMNL900911_AW.indd 79
ZMNL900911_AW.indd 79
• Switch the pump "ON"
• Immediately after the self-test the pump displays total volume delivered.
• Press the "CLR" key. The total volume delivered returns to "0 ml".
• The pump displays the previous programmed flow rate (ml/h).
• The pump is in the "HOLD" mode ready for programming.
BATTERY OPERATION
pump is equipped with a Lithium-ion battery. The expected battery life is approximate 2 to 5 years.
CONNECTION TO AN EXTERNAL ALARM SYSTEM
Nurse Call (corporate code 35752) enables connection to an external alarm system. This external
PDMS Cable (corporate code 35776) enables connection to an external PDMS system. Follow the
Infinity
PDMS Cable to connect the pump.
®
™
MAINTENANCE AND CLEANING
-79-
Infinity
Nurse Call to con-
®
™