The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user
®
of the Flocare
Infinity
enteral feeding pump should assure that it is used in such an environment.
®
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields
from RF
wireless
communications
equipment IEC
61000-4-3
Radiated RF
RTCA/DO-160E
Section 20
NOTE 1
At 80 Mhz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measurement field strength in the
location in which the Flocare
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the Flocare
Infinity
.
®
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
ZMNL900911_AW.indd 87
ZMNL900911_AW.indd 87
Guidance and manufacturer's Declaration – Electromagnetic immunity
IEC 60601 test level
Compliance level
3 Vrms
3 Vrms
150 KHz TO 80
MHz outside ISM and
amateur
radio bands
80% AM at 1 kHz
6 Vrms
6 Vrms
150 kHz to 80 MHz
in ISM and amateur
radio bands
80% AM at 1 kHz
10 V/m
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
9 V/m to 28 V/m
9 V/m to 28 V/m
15 specific frequencies
up to 5,785 GHz
75 V/m
No Equipment
100 MHz to 8 GHz
Category specified
(75 V/m)
Infinity
is used exceeds the applicable RF compliance level above, the Flocare
®
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Flocare
Infinity
, including cables, than the recommended sepa-
ration distance calculated from the equation applicable
toe the frequency of the transmitter:
Recommended separation distance
d = 1.2CD P
d = 1.2CD P
d = 2.3CD P
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,
less than the compliance level in each frequency range.
Interference may occur
in the vicinity of equipment
marked with the following
symbol:
-87-
®
80 MHz to 800 MHz
800 MHz to 6 GHz
should be
a
b
Infinity
should be observed
®
3.06.21 9:01