Accessories
Endocavity Balloon

INTENDED USE

The transducer standoff device can be used to perform scanning and needle guided procedures for body surface and endocavity diagnostic ultrasound.
CAUTION
Federal (United States) law restricts this device to sale by or on the order of a physician.
WARNING
Before use, you should be trained in ultrasonography. For instructions on the use of your transducer, see your system's user guide.
n
Do not reuse, reprocess or resterilize single-use device. Reuse, reprocessing or resterilization may create a risk of contamination of the device, cause
n
patient infection or cross-infection.
APPLYING BALLOON TO TRANSDUCER
1. Place an appropriate amount of gel inside endocavity balloon and/or on transducer face to aid in scan quality, application and removal of
balloon.
2. Insert transducer into balloon with fill tube oriented in top-center of transducer. Pull balloon tightly over transducer face to remove wrinkles and
air bubbles.
3. Visually orient balloon so lateral seams are symmetrical distance from longitudinal array.
4. Tape balloon securely making sure to attach around plastic tabs (Tape not included).
5. Fill syringe with 30cc of saline solution. Fill balloon through luer lock port holding prove tip down; fill and aspirate with 10-20cc until all air is
removed. Do not over inflate balloon.
6. Deflate balloon leaving 25-30cc syringe connected.
DISPOSAL
WARNING
Dispose of single-use components as infectious waste.