Cleaning Module-II
Installation
Direct
With this type of connection, the peristaltic pump of the Cleaning
Module-II is connected directly to the AMI transmitter and the pho-
connection to
tometer is connected to the peristaltic pump of the Cleaning Mod-
the AMI
ule-II.
transmitter
To connect the Cleaning Module-II to the AMI SAC254 proceed as
follows:
1 Switch off the analyzer according to chapter 6 «Stop of Opera-
2 Open the housing of the AMI transmitter. Disconnect the pho-
3 Feed the signal cable coming from the Cleaning Module-II into
4 Close the housing of the AMI transmitter.
5 Open the peristaltic pump of the Cleaning Module-II:
A-96.250.599 / 020620
WARNING
Electrical shock hazard!
Disconnect the AMI analyzer from the mains before connecting
the signal cable.
tion for Maintenance» in the corresponding manual.
tometer from the AMI transmitter and pull out the cable from the
cable gland.
the housing of the AMI transmitter. Connect the wires according
to the connecting diagram below.
–Release and remove the occlusion frames.
–Unscrew the 4 screws and remove the cover.
–Disconnect the motor connector [A].
A Motor connector
A