Cleaning Module-II
Maintenance
Dismount
pump tube
A-96.250.599 / 020620
The pump tubes can easily be dismounted and mounted. Proceed
as follows:
1 Remove the protection cap [G].
2 Open the occlusion frames [B] by turning them counter-clock-
wise.
3 Remove the pump tubes [D] from the rotor [C] by pulling the
complete occlusion frames [B] out of the holder.
4 Disconnect the reagent tubes from the old pump tubes and con-
nect them to the new pump tubes.
5 Install the new pump tubes by pushing the occlusion frames
onto the holder.
6 Lock the occlusion frames. Check that the occlusion frames and
the tubes are aligned perpendicular to the axis of the rotor.
7 Place the protection cap on the pump.
8 Insert the suction lances into the corresponding containers.
A
Pump housing
B
Occlusion frame open
A
C
Rotor
B
D
Pump tube
E
Pump inlet
C
F
Pump outlet
D
E
F