5.2

Replacing the fuses

When changing the fuses, always ensure that only fuses with the same values are used.
Replace the main fuse as follows:
1. Disconnect the device from the main power supply.
2. Apply a flat screwdriver to the slot at the bottom of the plug housing and gently pry it open.
3. Take the fuse holder out of the housing as far as possible and place it so that the fuses
point upwards.
4. Gently lift the black lever between the fuses and pull the fuse bracket out.
5. Remove and replace the fuses.
G
6. Put the fuse bracket back in the holder with the contacts pointing outwards.
7. Put the fuse holder back in the plug housing.
B
5.3
Warranty
Arthrex GmbH assumes a warranty to the first purchaser for a twelve month period with
regard to defects or failure of its instruments and electric devices. All defective products
covered by the warranty are repaired or replaced free of charge by Arthrex at Arthrex's
choice. The warranty does not cover any damage caused by unlawful use or improper
handling of the particular product.
The warranty becomes invalid when Arthrex products are changed in any way or repairs are
performed by anyone other than Arthrex.
Arthrex will gladly answer any questions referring to the quality, reliability and/or shelf life of
this product.
14
Operating Manual
Continuous Wave II Arthroscopy Pump
AR-6450