OBJ_DOKU-191483-003.fm Seite 17 Donnerstag, 9. Juni 2016 10:02 10
3.2

Check the product

Check the product before every occasion of use!
If damaged, do not use this product!
3.3

Assembling the product

Pull the protective sleeve over the ceramic tip of the APC appli-
cator. The outlet (1) of the APC applicator must be above the
midpoint of the protective sleeve and point away from it.
4

Cleaning, disinfection, sterilization

4.1

Safety instructions

Not suitable for manual cleaning and disinfection.
An ultrasonic bath can be used for precleaning.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.