3. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
4.9

Check

1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, burnt
edges.
If damaged, do not use this product!

4.10 Packaging

1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/polyethylene and/
or in a sterilization container.

4.11 Sterilization

Only sterilize products that have been cleaned and disinfected.
Erbe Elektromedizin recommends steam sterilization with the
method described below. If other sterilization methods are used,
Erbe Elektromedizin shall bear no responsibility.
Steam sterilization
• Pre-vacuum method with adequate product drying
• Holding time 3 to 18 minutes at 270 to 275 °F (132 to
135 °C) – with a tolerance in accordance with DIN EN
ISO 176655
• Sterilizers in accordance with applicable national standards
and regulations (e.g. DIN EN 13060 or DIN EN 285)
• Sterilization process validated according to DIN EN
ISO 17665
Please follow the recommendations issued by the sterilizer man-
ufacturer concerning loading, handling, etc.