3.2
Check the product
Before each use, check that the product is intact. Verify that
there are no signs of physical damage.
If damaged, do not use this product!
3.3
Assembling the product
Pull the protective sleeve over the ceramic tip of the APC appli-
cator. The outlet (1) of the APC applicator must be above the
midpoint of the protective sleeve and point away from it.
4
Cleaning, Disinfection, Sterilization
4.1
Safety Instructions
Not suitable for manual cleaning and disinfection.
An ultrasonic bath can be used for precleaning.
On no account use sharp or abrasive objects for cleaning purposes.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
During machine cleaning and disinfection, do not exceed 203 °F
(95 °C).
1