Reprocessing Limitation; Recommended Equipment / Substances; Recommended Methods; Dismantling - Erbe 20132-129 Manual Del Usuario

Lengueta protectora para orl
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
OBJ_DOKU-191483-003.fm Seite 18 Donnerstag, 9. Juni 2016 10:02 10
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
4.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
4.3

Recommended equipment / substances

The service life of the product, cleaning action and sterility are
validated for the following cleaning and preparation procedures
if the recommended or equivalent substances are used in accor-
dance with the manufacturer's instructions.
• Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning agent
neodisher ® mediclean forte (Dr. Weigert GmbH & Co. KG,
Hamburg, Germany) with the recommended washing program.
4.4

Recommended methods

Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
4.5

Dismantling

1. Remove the protective sleeve from the APC applicator.
18

Publicidad

Tabla de contenido
loading

Tabla de contenido