4.13 Modification of the product by the user
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
5
How to use
5.1
Checking the expiration date
The use-by-date is indicated on the packaging. Do not use this
product if the use-by-date has been exceeded!
5.2
Checking and opening the packaging
WARNING! Only use the product with intact, unopened sterile
packaging. You can identify packaging that is damaged, and
therefore no longer sterile from damaged seal seams, illegible
labeling, damp or even wet areas, for example.
Use aseptic techniques when removing the product from the
packaging!
5.3
Checking the product
CAUTION! Before use, check the insulation for damage, paying
particular attention to the ceramic insulation at the distal end of
the APC probe. Do not use this product if it is damaged.
5.4
Connecting the product and selecting unit settings
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the instrument.