8.2 Cleaning procedure

The following cleaning method has been validated according to the AAMI TIR 12
standard. Ambu makes no warranties concerning other cleaning and sterilisation
methods than the validated one described here. The validated method specifies
cleaning by washing machine using detergents.
Disassemble the devices prior to placing in the washer and program the washer
for the following parameters. Place devices into washer to allow for drainage:
Phase
Recirculation
time (minutes)
Pre-Wash 1
02:00
Enzyme wash 02:00
Wash 1
02:00
Rinse 1
0:15
Dry the devices appropriately.
Sterilisation by gravity steam autoclave running a full cycle at 134 degrees Celsius
for 10 minutes at a standard pressure and then dry 30 minutes.
12
Water
Detergent type and
temperature
concentration (if applicable)
Cold tap water
NA
Hot tap water
Enzymatic detergent
60.0 °C (140 °F)
Enzymatic detergent
(Set point)
Hot tap water
NA
(60.0 °C) (140 °F)
8.3 Parts that can be cleaned and sterilised
Applicable methods
X Applicable O
Machine Washing
Not applicable
Patient valve
X
Bag
X
Inlet valve
X
Extension tube
X
Silicone Facemask
X
Silicone Facemask size OA
X
Oxygen reservoir bag
X
Oxygen reservoir tube
X
8.4 Detergents and chemical disinfectants
Use only brands that are compatible with the resuscitator materials to avoid
reduction in the lifetime of the materials. Follow the instructions of the manufacturer
of the detergent or chemical disinfectant for dilution and exposure time.
CAUTION
Avoid using substances containing phenol to clean the product. Phenol will cause
premature wear and degradation of the materials or reduced product life span.
Promptly remove all residue of cleaning materials from the resuscitator. Residues
may cause premature wear or reduced product life span.
The following steps are generally recommended. Select proper methods for the
resuscitator parts in question according to the table above.
Disinfection & sterilisation
Autoclave 134°
X
X
X
X
X
0
X
0