The are two reasons for possible water leaks between
oxygenating module and holder D632:
1) the lever (ref. 8) is not in the "CLOSE" position;
2) during fixing of the oxygenator to the holder, lever (ref.
8) rotation was forced, causing breakage of the
oxygenator connectors.
In the first case turn the lever (ref. 8) to the "CLOSE"
position, in the second case replace the oxygenator.
F. REMOVAL OF OXYGENATOR FROM
HOLDER AT THE END OF BY-PASS
At the end of each procedure:
a) Turn the thermocirculator off.
b) Turn the lever (ref. 8) to the "OPEN" position.
c) Disconnect the water lines from the holder.
d) Remove oxygenator.
G. MAINTENANCE
Once a week lubricate the O-Rings (ref. 10) of the
Hansen female connectors of the water distributor with
silicon spray.
H. LIMITED WARRANTY
This Limited Warranty is in addition to any statutory rights of the
Purchaser pursuant to applicable law.
SORIN GROUP ITALIA warrants that all reasonable care has
been taken in the manufacture of this medical device, as
required by the nature of the device and the use for which the
device is intended.
SORIN GROUP ITALIA warrants that the medical device is
capable of functioning as indicated in the current instructions for
use when used in accordance with them by a qualified user and
before any expiry date indicated on the packaging.
However, SORIN GROUP ITALIA cannot guarantee that the
user will use the device correctly, nor that the incorrect
diagnosis or therapy and/or that the particular physical and
biological characteristics of an individual patient, do not affect
the performance and effectiveness of the device with damaging
consequences for the patient, even though the specified
instructions for use have been respected.
SORIN GROUP ITALIA, whilst emphasizing the need to adhere
strictly to the instructions for use and to adopt all the
precautions necessary for the correct use of the device, cannot
assume any responsibility for any loss, damage, expense,
incidents or consequences arising directly or indirectly from the
improper use of this device.
SORIN GROUP ITALIA undertakes to replace the medical
device in the event that it is defective at the time of placing on
the market or whilst being shipped by SORIN GROUP ITALIA
up to the time of delivery to the final user unless such defect
has been caused by mishandling by the purchaser.
The above replaces all other warranties explicit or implicit,
written or verbal, including warranties of merchantability and
fitness for purpose. No person, including any representative,
4
agent, dealer, distributor or intermediary of SORIN GROUP
ITALIA or any other industrial or commercial organization is
authorized to make any representation or warranty concerning
this medical device except as expressly stated herein. SORIN
GROUP ITALIA disclaims any warranty of merchantability and
any warranty of fitness for purpose with regard to this product
other than what is expressly stated herein. The purchaser
undertakes to comply with the terms of this Limited Warranty
and in particular agrees, in the event of a dispute or litigation
with SORIN GROUP ITALIA, not to make claims based on
alleged or proven changes or alterations made to this Limited
Warranty by any representative, agent, dealer, distributor or
other intermediary.
The existing relations between the parties to the contract (also
in the case that it is not drawn up in writing) to whom this
Warranty is given as well as every dispute related to it or in any
way connected to it as well as anything related to it or any
dispute concerning this Warranty, its interpretation and
execution, nothing excluded and/or reserved, are regulated
exclusively by the Italian law and jurisdiction. The court chosen
is the Court of Modena (Italy).
GB - ENGLISH

Publicidad

Tabla de contenido
loading

Tabla de contenido