ERROR INDICATORS AND TROUBLESHOOTING

ERROR
CAUSE
INDICATORS
Incomplete inflation of arm cuff
Cuff over-inflated and released
automatically
Could not obtain the pulse rate
Strong electromagnetic interference
(mobile phone or computer)
Irregular pulse rate or blood pressure
CORRECTION
• Check if the cuff is firmly connected to the device
• Make sure the cuff hose remains straight and unkinked during
measurement
• Reposition the cuff correctly (see pg. 16) and try the measurement again
• If the Er message persists, call product support
• Take a break, remove all clothing from upper arm, reposition the
cuff correctly (see pg. 16) and try the measurement again
• Make sure the cuff hose remains straight and unkinked during
measurement
• Reposition the cuff correctly (see pg. 16) and try the measurement again
• Please refrain from talking or moving during measurement
• Make sure the cuff hose remains straight and unkinked during
measurement
• Remove sources of interference (mobile phone or computer)
• Take a break, reposition the cuff correctly (see pg. 16) and try the
measurement again
• Take a break, reposition the cuff and your body position correctly
(see pg. 16) and try the measurement again
• Please refrain from talking or moving during measurement
21