Descargar Imprimir esta página

Automatic Cleaning; Inspection/Maintenance - CareFusion V. Mueller RH4701-001 Manual Del Usuario

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
areas. Note: It is recommended that the detergent solution
is changed when it becomes grossly contaminated (bloody
and/or turbid).
5. Rinse the device by completely immersing in tap water with
a temperature range of 27°C to 44°C (81°F to 111°F), for a
minimum of 30 seconds to remove any residual detergent
or debris.
6. Dry the device with a clean, lint-free towel.
7. Visually examine each device for cleanliness.
8. If visible soil remains, repeat cleaning procedure.

Automatic Cleaning

1. Ensure all pre-processing instructions are followed prior to
cleaning.
2. Clean the devices via the automatic cleaning parameters
below.
PHASE
RECIRCULATION
Pre-Wash
1
Enzyme
Wash
Wash 1
Rinse 1
Pure Rinse
Drying
3. If visible moisture is present dry the instrument with a
clean, lint-free towel.
4. Visually examine each instrument for cleanliness.
5. If visible soil remains, repeat cleaning procedure.

Inspection/Maintenance

Proper care and handling is essential for satisfactory
performance of any surgical device. The previous cautions
should be taken to ensure long and trouble-free service from all
your surgical devices. Inspect devices before each use for
broken, cracked, tarnished surfaces, movement of hinges, and
chipped or worn parts. If any of these conditions appear, do not
use the device. Return devices to an authorized repair service
center for repair or replacement.
Before sterilizing, lubricate the device with instrument milk or a
steam permeable/ water soluble lubricant, following the
lubricant manufacturer's instructions.
Let devices drip dry for three (3) minutes before packaging for
sterilization.
Ensure mirror is properly attached to the stainless steel shaft
and backing.
MINIMUM
TIME
15 Seconds
1 Minute
2 Minutes
15 Seconds
10 Seconds
N/A
WATER
TEMPERATURE
Cold Tap Water
1°C - 16°C
(33°F - 60°F)
Hot Tap Water
43°C - 82°C
(110°F - 179°F)
Tap Water
43°C - 82°C
(110°F - 179°F)
Tap Water
43°C - 82°C
(110°F - 179°F)
Purified Water
43°C - 82°C
(110°F - 179°F)
N/A
2
DETERGENT TYPE
AND CONCENTRATION
(IF APPLICABLE)
N/A
• Detergent: pH-
neutral/enzymatic
detergent
• Concentration: Per
the detergent
manufacturer's
recommendations
• Detergent: pH-neutral
cleanser
• Concentration: Per
the detergent
manufacturer's
recommendations
N/A
N/A
N/A

Publicidad

loading