S S e e t t u u p p
To unpack your product, see the unpacking instructions that are attached to the product inside of the shipping crate.
W W A A R R N N I I N N G G
• Always allow the product to reach room temperature before you set up the product or test functional operations.
Permanent product damage may occur.
• Always operate the product only when all operators are clear of the mechanisms.
• Always use mattress (6300-0-100) on the Stryker Model 6300 S S T T 1 1 and S S T T 1 1 - - X X Series stretcher. Use of any other
mattress may result in patient injury.
C C A A U U T T I I O O N N - - Always transport the S S T T 1 1 and S S T T 1 1 - - X X Series stretcher on floors made of wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the relative humidity should be at least 30% to avoid electrostatic discharge.
Make sure that the product works before you put the product into service.
1. Apply the brake. Push on the product to make sure that all four casters are locked.
2. Release the brake. Push on the product to make sure that all four casters are unlocked.
3. Raise and lower the litter with the hydraulic lift system.
4. Raise the product to the highest position and put the product in the Trendelenburg position. Make sure that the head end
lowers to the full down position.
5. Raise the product to the highest position and put the product in the Reverse Trendelenburg position. Make sure that the
foot end lowers to the full down position.
6. Apply the fifth wheel and make sure that the fifth wheel guides and pivots the product.
7. Make sure that the siderails raise, lower, and lock in place.
8. Raise and lower the manual Fowler backrest (head section).
KK-6300 Rev 00