P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the mattress before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
All product labels are in place and legible
All weldments (base frame, brake, litter, jack, carriage, IV pole pivot weldment, and push handle weldments) are not
damaged
Siderails move and latch
Siderail latches are secure
Siderail is not damaged
Siderail latch is not damaged, no burrs or debris in latch assembly
Antistatic caster is not worn or damaged
Casters lock with brake pedal applied
Casters are secure and swivel
Casters are free of wax and debris
Casters are not worn or damaged
Caster mounting joint is not damaged
Casters, brake mechanism, and brake rod are not damaged or cracked
Fowler raises, lowers, and latches in place
Fowler does not drift or drop unexpectedly
No leaks at the Fowler backrest cyclinders
Fowler gas cylinder pin is not stuck
Brake/steer pedals are not bent or damaged
Brake mechanism works
Steer function works
Fifth wheel is not worn or damaged and works
Fifth wheel linkage is not bent or overtraveled
No debris or wax buildup in fifth wheel
Carriage bolt is secure
Base frame is not damaged
Pump pedal is not loose, worn, or damaged
Hydraulic release pedals are not loose or damaged
Jack release valve is free of dust, debris, does not stick
Jack linkages are not out of adjustment or damaged
Jack adjustment valves and spring work
Jacks are not damaged
Head end and foot end jacks raise and lower at the same time