5.9.5 Error Messages, Errors

E1
E2
E3
E4
E5
E6
E7
E12
E18
E19
E26
E28
E29
E30
E31
E36
E37
E38
E39
774 Oven Sample Processor, Instrucciones para el uso
Detailed Status Conditions
Status conditions of the global $R:
Basic state: ready to start automatic processing
An assembly step has been executed.
Status conditions of the global $G:
Instrument at the beginning of processing
Instrument processing the start sequence, displays line number and
current command
Instrument processing the sample sequence, displays line number and
current command
Instrument processing the final sequence, displays line number and
current command
Instrument processing a manual command
Instrument processing an assembly-command
Status conditions of the global $H:
The status conditions of the global $H are identical with the ones of the
global $G.
Fatal Instrument Errors:
Incorrect program check sum
RAM read/write error
RAM lost data
Timer interrupt for multi-tasking missing
RS232 module test error
RS232 read/write error
Display read/write error
EBUS error
Low battery
RAM test error
Program-specific Messages:
Manual stop
Wrong object call
Wrong value or no value allowed
Wrong trigger
Comman currently not allowed
RS Receive Errors:
Parity error
Exit: <QUIT> and set the same parity for both instruments.
Stop bit error
Exit: <QUIT>and set the same stop bit for both instruments.
Overrun error. At least 1 character could not be read.
Exit: <QUIT>.
Internal receive buffer full (>82 characters).
Exit: <QUIT>.