Parameter List; Setup"/"Balance" Menu - Sartorius Entris BCE6202 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 131
System Settings
7.3

Parameter List

7.3.1

"SETUP"/"BALANCE" Menu

Parameter
Setting values
AMBIENT
V .STABLE
STABLE*
UNSTABL .
V .UNSTBL .
APP FILT .
FINAL .RD .*
FILLING
STABIL .
V . ACC .
ACC .*
FAST
ZER ./TAR .
W/O STB .
W/ STAB .*
AUTOZER .
ON*
OFF
* Factory setting
18
Entris
®
BCE Operating Instructions
Explanation
Sets the ambient conditions to "very stable": Activates a fast change in the weight values in
the event of a load change with a high output rate .
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
Sets the ambient conditions to "stable" . Recommended for the following work environment:
− Stable table
− Slight movement in the room
− Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in weight values with
a reduced output rate . Recommended for the following work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly delayed change in the
weight values and long wait for stability with a further reduction in the output rate .
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
Activates a filter that enables a fast change in the display for very fast load changes .
Display changes with minimal load changes (in the digit range) occur more slowly .
Activates a filter that enables a very fast change in the display with minimal load changes,
e .g . when filling containers .
Sets the stability to "very accurate" .
Sets the stability to "accurate" .
Sets the stability to "fast" .
If the button is pressed: The function of the [Zero] or [Tare] button is executed immediately .
The function of the [Zero] or [Tare] button is only executed after stability is achieved .
Activates automatic zeroing . The display is automatically set to zero in case of a deviation of
0 less than (X) .
Deactivates automatic zeroing . Zeroing must be triggered with the [Zero] button .

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido