Malfunctions
10.2 Troubleshooting
Fault
The operating display is
blank .
The displayed weight value
changes constantly .
The weight readout
displayed by the device
is obviously wrong .
For a conformity-assessed
device: The [Invalid weight
value] display appears .
32
Entris
®
BCE Operating Instructions
Cause
The device is disconnected .
The power supply unit is not
connected .
The installation site is unstable .
A foreign object is positioned
between the weighing pan and
the housing .
The device has not been calibrated
and adjusted .
The device was not tared before
weighing .
The cause of this fault is displayed
after pressing the [Change] key . At
the same time, the "Toggle between
weight units" function is locked .
ISOCAL .E .: The device needs to be
calibrated and adjusted .
WARMU .xx .: The device is in the
warm-up phase and has not yet
reached its operating temperature .
xx = remaining time in minutes
VALUE .ERR .: The displayed value is
invalid .
Remedy
Check the connection to the power
supply .
Connect the power supply cable to the
power supply .
Adjust the parameter for the ambient
temperatures .
Change the installation site .
Remove the foreign object .
Calibrate and adjust the device .
Tare the device .
Calibrate and adjust the device .
Comply with the warm-up time after
switching the device on .
Set the device to zero .
Chapter, page
8 .4, 27
8 .4, 27
15 .7, 37