Technical Data
15.10.3 Models BCE822 | BCE622
Model
Readability | Scale interval (d)
Maximum capacity (Max)
Repeatability
At 5% load, typical value
At approx . the maximum load, typical value
Linearity deviation
Limits
Typical value
Sensitivity drift between +10 °C and +30 °C
Tare maximum capacity (subtractive)
isoCAL (only for i-1x models):
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weighing according to USP (United States
Pharmacopeia), Chap . 41
Optimum minimum initial weighing
Typical measurement time
Typical stabilization time
Recommended Calibration Weight
External calibrated test weight
Accuracy class in accordance with OIML R111-1
Weighing pan size
Net weight, approx .
40
Entris
®
BCE Operating Instructions
Unit
Value
Value
mg
10
10
g
820
820
mg
5
5
mg
10
10
± mg
20
20
± mg
6
6
± ppm/K 2
3 .5
<100% of maximum capacity
K
2
h
6
II
BC_EE
mg
100
mg
500
g
8 .2
8 .2
s
<1 .0
<1 .0
s
<0 .9
<0 .9
g
500
500
F2
F2
mm
182 x 182
182 x 182
kg
4 .3 | 6 .2
4 .3
Value
Value
10
10
620
620
5
5
10
10
20
20
6
6
2
3 .5
2
6
II
BC_EE
100
500
8 .2
8 .2
<1 .0
<1 .0
<0 .9
<0 .9
500
500
F2
F2
182 x 182
182 x 182
4 .3 | 6 .2
4 .3