Alibi Memory - RADWAG R Balances Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

The data that is to be printed out, depends on the settings of parameter P5.3 GLP PRINTOUT.
Parameters selected for the printout (<YES> option), are printed out also on measurement result
printout for WEIGHINGS database (see sec. 8.2.)

9.5. ALIBI MEMORY

The balance is equipped with "ALIBI", a type of memory that allows you to save and store up to
100 000 measurements carried out on the balance.
If the "ALIBI" memory is installed in the balance, the saving of the measurements occurs
automatically, by pressing <PRINT> button, without a need of additional actions or settings change.
The additional data is saved along with the measurement.
Date of the measurement
Time of the measurement
Measurement result (mass)
Tare value
Person carrying out the measurement (logged-in user)
The name of the product that has been weighed
The software saves the measurements in a so called loop, i.e. when the measurement 100 001 is
saved, the measurement 1 is automatically deleted from the balance's memory.
The measurements saved in the balance's memory cannot be deleted.
It is possible for a user to view and print out the data saved in the "ALIBI" memory.
Procedure:
Each measurement is saved with its individual number. The format is: b4.5.n, where <n> is the
consecutive number of the saved measurement. In the bottom line the date and time are displayed
for every single measurement.
Operations for ALIBI database and WEIGHINGS database are likewise, for detailed information read
the previous section.
An exemplary printout:
Date
Time
Result
Tare
User
Product
To return to the weighing mode press
19.06.2013
6:48:41
199.90 g
0.000 g
SMITH
PILL
button repeatedly.
- 34 -