RADWAG R Balances Serie Manual Del Usuario página 90

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Value release
Format: SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command's function is similar to function of PRINT button located on an overlay,
Internal adjustment
Format: IC CR LF
Response options:
IC_A CR LF
- command understood and in progress
IC_D CR LF
- calibration completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for a stable measurement result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal adjustment of the balance
Format: IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
IC1_OK CR LF
- command carried out
Operation is disabled for verified balances.
For non-verified balances the command inhibits internal calibration until it is enabled via IC0
command or until the balance is turned off. The command does not modify settings specifying
calibration start.
Enable automatic internal adjustment of the balance
Format: IC0 CR LF
Response options:
IC0_I CR LF
- command understood but not accessible at this moment
IC0_OK CR LF
- command carried out
Operation is disabled for verified balances.
Give balance serial number
Format: NB CR LF
Response options:
NB_A_"x" CR LF - command understood, response: serial number
NB_I CR LF
- command understood but not accessible at this moment
x – serial number of the device (inserted in between inverted commas)
An example:
Command:
NB CR LF – return serial number
Response:
NB_A_"1234567" – serial number of the device – "1234567"
Lock balance keypad
Format: K1 CR LF
Response options:
K1_I CR LF
- command understood but not accessible at this moment
K1_OK CR LF
- command carried out
Command locks the balance keypad (proximity sensors, touch panel) until the moment of turning the
balance off or until sending K0 command
Unlock balance keypad
Format: K0 CR LF
Response options:
K0_I CR LF
- command understood but not accessible at this moment
K0_OK CR LF
- command carried out
Give available working modes
Command overview: Command returns accessible working modes.
Format: OMI <CR><LF>
Response options:
- 90 -

Publicidad

Tabla de contenido
loading