Setting The Reference Mass: Entering Mass Value; Animal Weighing; Additional Settings - RADWAG R Balances Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

11.6.2. Setting the Reference Mass: Entering Mass Value

Procedure:
Press F button to which function <ENTER SAMPLE> is assigned, <ENTER SAMPLE> editing
window is displayed.
Use arrow buttons to enter the known reference mass value, press
The software automatically enters <PERCENT WEIGHING> mode displaying the value of
0.000%, the bottom line displays entered reference mass value (if such an option has been
selected for <INFORMATION> function).

11.7. ANIMAL WEIGHING

Animal Weighing mode enables reliable determining of mass of objects that is in motion. In principle,
objects in motion generate unstable measurement, thus Animal Weighing mode requires using a
different filtering method of measurement signal.

11.7.1. Additional Settings

Apart from standard settings for this mode (i.e. weighing mode settings) some additional parameters
have been designed to describe operation of the mode.
Additional settings list:
AVERAGING TIME – Amount of time during which recorded measurement results are
analysed. Obtained data is used to determine average measurement result.
THRESHOLD - It is a value expressed in mass measuring units. In order to start measurement,
the indication value has to exceed the set threshold value.
AUTOSTART - Determines start criterion for measurements: whether they are to be initiated
manually,
on
When the parameter is set to <YES> option, the measurement starts automatically if an
indication on a balance display exceeds the set threshold value. The following measurement
can start on removing the weighed object from a balance pan (the indication has to return
below the value set in the threshold), and on loading the pan with a new object at a moment
when the set threshold value indication is exceeded.
Prior to tests it is necessary to adjust settings of the above options so that correct values were
provided. The selected options shall be respective to needs and expectations resulting from working
environment.
pressing
a
button
or
selecting
- 54 -
button to confirm.
START,
or
automatically.

Publicidad

Tabla de contenido
loading

Tabla de contenido