Set the ambient humidity, next press ENTER for confirmation. The computer software proceeds
to the next step, i.e. air pressure settings.
Set the air pressure, next press ENTER for confirmation. The computer software proceeds to
the next step, i.e. entering the first control volume (V1) for a tested pipette.
Enter (V1) control volume. For fixed volume pipettes (P2.13.5 VOLUME DETERMINATION
parameter set to value <1>) this is the one and only value to be entered. The computer
software proceeds to the next step, i.e. entering the second control volume (V2) for a tested
pipette.
Enter (V2) control volume. The computer software proceeds to the next step, i.e. entering the
third control volume (V3) for a tested pipette.
Enter (V3) control volume.
Upon entering the data the computer software proceeds to pipettes calibration.
Follow the displayed description and complete the procedure.
By means of a pipette, dose the first portion of water, wait until the measurement is stable and
press ENTER to confirm.
The computer software records the measurement. If P2.3.7 AUTOMATING TARING parameter
is set to value <YES> then the indication is automatically tared. If P2.3.7 AUTOMATIC
TARING parameter is set to value <NO> than the user must press TARE button in order to tare
the indication before dosing another portion of water.
Upon completion of series of measurements, summary is displayed.
Press ENTER to continue measuring process, press Esc. to stop it. Upon pressing Esc. button
the computer software returns to the main window.