I. INTRODUCTION
I. 1. Description of the Device
OSIRIS 2 is an adjustable, controlled frequency ventilator,
with an adjustable I/E ratio from 1/1 to 1/3 by steps of 0.5.
It ensures controlled volume ventilation with either:
• pure oxygen (VAC/ACMV 100 % O
• or with a mixture of O
(VAC/ACMV Air + O
It also ensures spontanteous ventilation, pressure assisted
(VS-AI/PS) which may be made non-invasively.
The PEEP* function is integrated and adjustable from 0 to
15 mbar (0 to 15 cmH
OSIRIS 2 has audible and visual alarms to signal any
ventilation faults. The setting of a maximum pressure
threshold (Pmax) fixes a pressure limit and therefore
avoids any barotrauma (insufflation stops as soon as the
set threshold is passed, and the audible and visual alarms
are activated).
PEEP* : Positive End Expiratory Pressure
I. 2. Intended Use
OSIRIS 2 is intended for use by hospital personnel
(physicians, nurses, etc.) and non-hospital emergency
personnel. It is used:
• For transportation between hospitals,
• For emergency transport between hospitals,
• In post-operative recovery rooms,
• In intensive care units,
• For resuscitation.
For post-operatory or intensive care use, OSIRIS 2 must
have currently approved monitoring equipment.
I. 3. Composition of OSIRIS 2
• The basic OSIRIS 2 unit .
• Three complete single-use patient circuits.
• A 3 meter O
• One adult mask.
• One child's mask.
• A mains supply.
• A carrying case.
II. DESCRIPTION AND SETTINGS
II. 1. FRONT PANEL

Identification of parts

Display screen - parameters, settings,
measurements. 2-line display.
Ventilation mode switch
This switch allows the unit to be made operational. It must
be switched to one of the 3 positions: VAC/ACMV 100%
or VAC/ACMV Air + O
To stop the unit, turn the switch to position 0.
PEEP setting
This knob allows the PEEP to be set, from 0 to 15 cmH
A safety device
extended range of 10 to 15 cmH
YL024400 - V 2.1 - 10/2012
),
2
/ Ambiant air (FiO
2
2
).
2
O).
2
supply tube.
2
or VS-AI/PS.
2
must be released to give access to the
O.
2
EN
= about60 %)
O.
2
3